MEASUREMENT AND UNCERTAINTY OF THE SIZE-OF-SOURCE EFFECT OF RADIATION THERMOMETERS, IN THE CONTEXT OF AN INDUSTRIAL CALIBRATION LABORATORY
Keywords:
Radiation thermometer, thermometry, industrial calibration, metrology, traceability, blackbody, emissivityAbstract
An experimental procedure for characterizing the size-of-source
effect is proposed. It is one of the main influence variables for
uncertainty, in both calibration and use of direct reading radiation
thermometers. The procedure and uncertainty calculation
described are aligned, in terms of metrological traceability, with the
requirements generally imposed to ensure the accuracy of
measurements in industry and science. Finally, results of
application and validation of the procedure, with equipment
normally found in radiation thermometry calibration laboratories, in
the industrial field, are shown.